I. Basic Setting Once the instrument installation is complete, simply set the installation height and select the standard weir/flume type for operation. II. Installation Height Setting Depending on the type of water measuring weir/flume, the installation height varies. Below are the methods to confirm the installation height for standard water measuring weirs/flumes: Parshall Flume: Measure the vertical distance from the ultrasonic transducer (probe) end to the bottom inside the water measuring flume, in meters. Rectangular Weir: Measure the vertical distance from the ultrasonic transducer (probe) end to the lowest point of the water outlet, in meters. Triangular Weir: Measure the vertical distance from the ultrasonic transducer (probe) end to the right-angle point of the water outlet, in meters.
III. Flume/Weir Selection and Geometric Parameters Setting Choose the standard flume/weir type based on the actual type. Set the corresponding geometric parameters for each type. Below are the geometric parameters for common flumes/weirs: Parshall Flume: Select 'Parshall Flume' and designate the standard flume number. The width of the throat section determines the flume number, as shown in Table 1. Rectangular Weir: Select 'Rectangular Weir' and set parameters such as channel width, weir width, bottom height, and a description of the geometric parameters.